Gas Analysis
Gas analysis
Controlling the accuracy of an incubator in regulating CO2 and O2 is one of the standard procedures in IVF laboratories. The K-Systems CO2& O2 analyzer is a semi-automatic device making your routine checks easy and reproducible. It is easy to calibrate using outside air and 100 % CO2.
Specifications:
Min 2,500 measurements with one battery load (four alkaline batteries, type AA)
Memory function for 10 measurements
Standard measuring time 10 sec,
Approx. 15 ml gas volume per measurement
Calibration gas: Outside air or 0 % CO2 and 100 % CO2
O2 sensor: electrochemical; resolution 0.1 %; accuracy 0.25 % abs. ± 2 % rel. if CO2 < 25 %, 0.5 % abs. ± 2 % rel. if CO2 ≥ 25 %; expected life time > 1 year
CO2 sensor: non-dispersive infra-red, temperature compensated; resolution 0.1 %, Accuracy:
2 % abs. at 0 - 20 %, 3 % abs. at 20 - 100 %; expected life time > 3 years
Weight 0.5 kg
Dimensions 74 x 63 x 135 mm (W x H x D)
Advantages:
- Robust and ergonomically designed aluminium casing
- One device – two analytes: CO2 and O2
- Auto-stop function prevents sensor from being damaged
- Memory function
Ref. No. 11092
K-Systems O2 & CO2 analyzer cpl. with sampling tube, manual, batteries in a black plastic case